差分脉冲阳极溶出伏安法检测重金属离子
[Abstract]:A new differential pulse anodic stripping voltammetry was used to detect heavy metal ions of lead and cadmium in liquid environment. By changing different experimental conditions and optimizing experimental parameters, lead and cadmium ions in liquid phase were determined by co-plating mercury with glassy carbon electrode. The detection limit for lead and cadmium is 0.54 ~ 0.79 g / L, respectively, and the linear correlation coefficient of cadmium ion is 0.999 7. The differential pulse anodic stripping voltammetry has the advantages of high sensitivity, simple instrument and equipment, simple operation, small pollution, and can be measured repeatedly. It is an efficient and cheap method to detect the concentration of heavy metals. It lays a foundation for the real-time monitoring of heavy metal ions.
【作者单位】: 成都信息工程大学光电技术学院;四川农业大学理学院;
【基金】:国家自然科学基金(61701050)