Mallinckrodt高压注射器故障维修一例

发布时间:2018-08-19 16:23
【摘要】:正故障现象高压注射器型号Angiomat IIIumena,开机后放入针筒,关上注射筒门并推动加注控制条排气,设置参数准备注射,Configuration屏幕上出现提示:injector disabled,faceplate open,press main screen key to continue。注射头屏幕见图1。按[MAIN SCREEN]键重新开始启动过程,故障依然如前,注射失败。
[Abstract]:Positive fault high pressure syringe model Angiomat IIIumena, put into the syringe after boot, close the injection tube door and push the injection control strip exhaust, set the parameters to prepare the injection configuration screen with the prompt: injector disabled plate openplate press main screen key to continue. The injection head screen is shown in figure 1. Press the [MAIN SCREEN] button to restart the process, the fault is still as before, injection failed.