当前位置:主页 > 科技论文 > 电子信息论文 >

用于探测器中红外绝对光谱响应度测量的激光源

发布时间:2018-07-20 13:07
【摘要】:为了实现中红外探测器绝对光谱响应度高准确度的测量,从理论和实验上研究了中红外波段激光的功率稳定性的提高和光束质量的优化方法。采用声光调制和反馈控制的方法,把中红外激光的功率稳定性提高到0.1%以内;根据高斯光束传输理论计算了所需空间滤波器的各项参数,设计了中红外空间滤波器,搭建了相应的实验装置,显著提高了中红外激光光束质量。为中红外绝对光谱响应度的高准确度测量提供了一个可靠的中红外波段激光光源。
[Abstract]:In order to measure the absolute spectral responsivity of the mid-infrared detector with high accuracy, the methods of improving the power stability and optimizing the beam quality of the mid-infrared laser are studied theoretically and experimentally. By using acousto-optic modulation and feedback control, the power stability of the mid-infrared laser is improved to less than 0.1%, and the parameters of the required spatial filter are calculated according to the Gao Si beam propagation theory, and the mid-infrared spatial filter is designed. A corresponding experimental device has been set up to improve the beam quality of mid-infrared laser. A reliable mid-infrared laser source is provided for the high accuracy measurement of the absolute spectral responsivity.
【作者单位】: 中国电子科技集团公司第四十一研究所;电子测试技术国防科技重点实验室;
【基金】:国家重点基础研究发展计划(863计划) 空间辐射测量基准源研制基金(2015AA123702) 国防技术基础项目(J3120BA001)
【分类号】:TN215

【相似文献】

相关期刊论文 前3条

1 史学舜;陈坤峰;王恒飞;孙权社;;硅陷阱探测器绝对光谱响应率校准实验研究[J];光学与光电技术;2009年05期

2 Michael A. Lind ,缪家鼎,戴君武;探测器的绝对光谱响应的测量[J];半导体光电;1981年04期

3 ;[J];;年期

相关会议论文 前1条

1 全治科;高业胜;王恒飞;;基于低温辐射计的探测器绝对光谱响应率测量[A];第十二届全国光学测试学术讨论会论文(摘要集)[C];2008年