自含式压力蒸汽灭菌生物指示剂D值测定的影响因素
发布时间:2018-06-13 14:26
本文选题:自含式生物指示剂 + 嗜热脂肪芽孢杆菌 ; 参考:《中国消毒学杂志》2017年07期
【摘要】:目的 研究自含式压力蒸汽灭菌生物指示剂D值测定的影响因素,提高D值测定的准确性。方法 使用生物指示剂抗力测定仪,对不同生产企业的自含式压力蒸汽灭菌生物指示剂进行D值测定,比较透气孔、培养基、放置方式、载样器材对D值测定的影响。结果 透气孔、培养基、放置方式、载样器材对D值的最大影响率分别为6.9%、17.3%、29.9%、38.2%。结论 放置方式和载样器材是影响自含式压力蒸汽灭菌生物指示剂D值测定的主要因素,建议用带孔支架以水平方式进行测定。
[Abstract]:Objective to study the factors influencing the determination of D value of self contained autoclaving biological indicator and to improve the accuracy of D value determination. Methods using biological indicator resistance tester, the D value of self-contained autoclaving biological indicator of autogenous steam sterilization in different production enterprises was determined, and the effects of air permeability, culture medium, placement method and sample loading equipment on D value were compared. Results the maximum influence rates of air permeation hole, culture medium, placement method and sample loading equipment on D value were 6.9% and 17.3% and 29.9% and 38.2% respectively. Conclusion the method of placing and loading equipment are the main factors that influence the determination of D value of self-contained autoclaving biological indicator. It is suggested that the porous scaffold should be used to measure the D value of the biological indicator.
【作者单位】: 四川省食品药品检验检测院;四川科伦药业股份有限公司;
【分类号】:R187
【相似文献】
相关会议论文 前4条
1 詹淼;马桂玲;;快速阅读生物指示剂监测灭菌效果在植入物中的应用观察[A];全国第四届医院消毒供应中心护理学术交流暨专题讲座会议论文汇编[C];2008年
2 詹淼;张红玲;王晓东;;快速阅读生物指示剂监测灭菌效果在植入物中的应用观察[A];消毒供应中心(室)建设与管理高级研修班暨学术会论文集(1)[C];2008年
3 周莉;秦红梅;;自动阅读器监测生物指示剂的使用及结果分析[A];全国口腔科护理学术交流暨专题讲座会议论文汇编[C];2005年
4 褚金萍;;1292快速生物监测指示剂的应用体会[A];中华护理学会第5届全国消毒供应中心发展论坛论文汇编[C];2009年
相关硕士学位论文 前1条
1 王新建;最终灭菌药品无菌保证体系和参数放行的关键控制因素探讨[D];山东大学;2006年
,本文编号:2014322