不同方法干燥管腔器械的效果比较
发布时间:2018-07-09 21:19
本文选题:管腔器械 + 气枪 ; 参考:《中国消毒学杂志》2017年02期
【摘要】:目的比较不同方法干燥管腔器械的效果。方法将300件管腔器械(包括腔镜器械、超声刀和气腹管)分为3组,每组100件,完成终末清洗后,在温度恒定条件下,用气枪或干布将器械表面水分擦去,干燥时间均为25 min。A组采用温度设为70℃普通医用高温干燥柜干燥;B组采用温度设为70℃普通医用高温干燥柜加高压气枪吹干燥;C组采用新型低温真空干燥柜干燥。运用目测法和物理方法比较3种方法的干燥效果。结果 A组36%管腔器械干燥后有少量残留水份;B组18%管腔器械干燥后有少量的残留水份;C组1%管腔器械干燥后有少量残留水份。结论 C组采用新型的低温真空干燥技术优于A、B组传统的干燥柜技术,值得临床推广。
[Abstract]:Objective to compare the effects of different methods for drying lumen instruments. Methods 300 lumen instruments (including endoscopic instruments, ultrasonic knives and pneumoperitoneal tubes) were divided into 3 groups, 100 in each group. After finishing the final cleaning, the surface moisture of the instruments was erased with air gun or dry cloth under constant temperature. The drying time was 25 min.A. The drying temperature of medical high temperature drying cabinet was 70 鈩,
本文编号:2110690
本文链接:https://www.wllwen.com/yixuelunwen/swyx/2110690.html