宝石能谱CT低电压联合自适应性统计迭代重建在腹部扫描中的应用
本文选题:低电压 + 迭代重建 ; 参考:《山东医药》2015年35期
【摘要】:目的观察宝石能谱CT低电压联合自适应性统计迭代重建在腹部扫描中的应用价值及对图像质量的影响。方法 200例行上腹部增强检查的患者随机分为A、B两组各100例,A组采用低电压(100 k V),并进行自适应性统计迭代重建40%重建。B组采用常规电压(120 k V),并进行滤波反投影(FBP)重建;另根据BMI(23、23~28、28)将每组分为3个亚组(A组分为A1、A2、A3组,B组分为B1、B2、B3组)。记录各亚组剂量报告中CT剂量容积指数(CTDIvol)及剂量长度乘积(DLP),并计算有效剂量(ED);测量腹主动脉、门静脉血管腔内的CT值,并计算以上血管的信号噪声比(SNR)和对比噪声比(CNR);从肠系膜动脉的清晰度、门脉分支的清晰度、图像整体质量、主观噪声评分四个方面对图像进行主观评价。结果与B2组比较,A2组CTDIvol、DLP、ED减小(P均0.05);与B3组比较,CTDIvol、DLP、ED减小(P均0.05)。与B1组比较,A1组主动脉CT值、门静脉CT值、门静脉CNR增大(P均0.05);与B2组比较,A2组主动脉CT值、主动脉CNR、门静脉CT值增大(P均0.05);与B3组比较,主动脉CT值、门静脉CT值增大(P均0.05)。A1、B1组肠系膜血管清晰度、门静脉分支的清晰度、图像整体质量、主观噪声评分比较,P均0.05,其各自相应秩和检验量为486、475、502.5、370.75;A2、B2组肠系膜血管清晰度、门静脉分支的清晰度、图像整体质量、主观噪声评分比较,P均0.05,其各自相应秩和检验量为1 316、1 316、1 385.5、1 223;A3、B3组肠系膜血管清晰度、门静脉分支的清晰度、图像整体质量、主观噪声评分比较,P均0.05,其各自相应秩和检验量为140、130、161、118。结论低电压腹部扫描能明显降低辐射剂量,提高血管内碘浓度,结合自适应性统计迭代重建,可以降低图像噪声,改善图像质量,满足临床诊断要求。
[Abstract]:Objective to observe the application value of low voltage combined adaptive statistical iterative reconstruction of gemstone energy dispersive CT in abdominal scanning and its effect on image quality. Methods 200 patients with epigastric enhancement were randomly divided into two groups: group A (n = 100) and group A (n = 100). In addition, each group was divided into three subgroups: group A, group A, group A, group A, group A, group A, group A, group B, group B, group B, group B, group B, group B, group B, group B, group B, group B, group B, group B, group B, group B, group B, group B, group The CT dose volume index (CTDIvoll) and dose length product (DLPN) were recorded in each subgroup dose report, and the effective dose EDV was calculated, and the CT values of abdominal aorta and portal vein were measured. The signal noise ratio (SNR) and contrast noise ratio (CNR) of the above vessels were calculated, and the image was evaluated subjectively from four aspects: the definition of mesenteric artery, the clarity of portal branch, the overall image quality, and the subjective noise score. Results compared with group B2, the Ed of CTDIvolo DLPnD in group A _ 2 decreased by 0.05 and that in group B _ 3 decreased by 0.05. Compared with group B1, CT value of aorta, CT value of portal vein and CNR of portal vein were 0.05 in group A1, 0.05 in group A 2, 0.05 in group A and 0.05 in group B 2, respectively, and in group B 3, CT value of aorta was higher than that in group B 3, the CT value of aorta in group A 2 was higher than that in group B 2, and the CT value in group B 2 was higher than that in group B 2. Portal CT increased mesenteric vascular clarity, resolution of portal vein branches, overall image quality, subjective noise score (P < 0.05), and the corresponding rank sum test was 486475502.5n370.75A _ 2 / B _ 2 in group B _ 2 of mesenteric blood vessel clarity, and the results showed that the resolution of mesenteric vessels in group B _ 2 was significantly higher than that in group B _ 2 (P < 0.05). The definition, overall image quality and subjective noise score of portal vein branches were all 0.05. The corresponding rank sum test of portal vein branches was 1 316 / 1 3165 / 1 223A 3 / B 3 group respectively, and the definition of portal vein branches and the overall image quality of each group were compared. The scores of subjective noise were 0.05, and the corresponding rank sum test was 1401301 161118. Conclusion low voltage abdominal scan can significantly reduce radiation dose, increase intravascular iodine concentration, combined with adaptive statistical iterative reconstruction, can reduce image noise, improve image quality, and meet the requirements of clinical diagnosis.
【作者单位】: 鄂尔多斯市中心医院;
【分类号】:R816.5
【参考文献】
相关期刊论文 前2条
1 杨尚文;何健;杨献峰;周科峰;朱斌;;ASiR算法结合自动管电流调制技术在胸部低剂量CT中的应用研究[J];医学影像学杂志;2012年01期
2 吴瑶媛;王万勤;刘斌;Isao Tanaka;张帅;;FBP、ASiR和VEO三种重建算法对常规剂量胸部CT图像质量的影响[J];中国医学影像技术;2012年03期
【共引文献】
相关期刊论文 前10条
1 刘婧;王霄英;许玉峰;邱建星;曹会志;胡晓煜;蒋学祥;;迭代重建在前列腺CT成像中的应用研究[J];放射学实践;2013年03期
2 李莉;周恩利;刘娜;张宝霞;;低剂量CT扫描在外伤性肋骨骨折检查中的应用[J];国际医学放射学杂志;2014年01期
3 沈合松;王馨华;罗明月;梁丹;朱珊珊;朱冬云;杜彩虹;;640层CT冠状动脉造影AIDR3D重建算法的图像质量及辐射剂量评价[J];重庆医学;2015年10期
4 杨蕾;相爱华;邓昆;杨瑞山;;迭代重建技术在育龄人群泌尿系结石CT检查中的初步应用[J];中国医药科学;2015年11期
5 孙记航;彭芸;张祺丰;于彤;王冬潜;刘勇;蒋玲;;儿童低剂量胸部CT中自适应迭代重建权重最佳阈值的初步研究[J];临床放射学杂志;2014年04期
6 浦仁旺;刘义军;刘静红;卢绪论;王诗瑜;刘爱连;;低管电压结合自适应统计性迭代算法对腹部CT平扫图像的影响[J];中国临床医学影像杂志;2015年01期
7 谢鹏;查云飞;姜伦;陈顺新;邵天喜;;基于ASIR技术的低管电压MDCT肾动脉成像的可行性研究[J];临床放射学杂志;2015年01期
8 顾金凤;刘斌;;64排螺旋CT胸部低剂量扫描方案的研究[J];中华疾病控制杂志;2015年06期
9 刘静红;刘爱连;浦仁旺;吕婷婷;田士峰;李烨;刘晓峰;刘义军;;能谱CT低电压低对比剂浓度结合ASIR技术对胆囊动脉图像质量影响的研究[J];临床放射学杂志;2015年08期
10 曲岷;;适应性统计迭代重建技术在低剂量CT扫描中的应用进展[J];实用医学影像杂志;2014年06期
相关硕士学位论文 前9条
1 刘婉如;应用低剂量扫描技术与低浓度对比剂行宝石CT增强扫描的实验研究[D];吉林大学;2013年
2 钟欢欢;640层容积CT低剂量冠状动脉成像的实验研究[D];遵义医学院;2014年
3 左紫薇;最佳ASIR权重低剂量扫描在胸主动脉CTA中的应用[D];河北大学;2014年
4 杨石平;双源CT颞骨HRCT低剂量扫描及轻度先天感音神经性耳聋内耳骨性结构测量[D];昆明医科大学;2014年
5 左倩;利用迭代重建和能谱技术的CT结肠成像实验研究[D];天津医科大学;2014年
6 许品;宝石CT低剂量扫描结合低浓度对比剂行胸痛三联检查的临床应用研究[D];大连医科大学;2014年
7 马宇;ASiR算法结合自动管电流调制技术在腹部CT低剂量的应用研究[D];桂林医学院;2014年
8 杨富阁;东芝320排低剂量CT灌注成像及AIDR重建技术在直肠癌中的应用[D];郑州大学;2014年
9 王杰;双源CT胸部扫描迭代重建算法与图像质量及辐射剂量相关性的研究[D];重庆医科大学;2014年
【二级参考文献】
相关期刊论文 前2条
1 张翼;赵斌;;低剂量螺旋CT扫描技术[J];医学影像学杂志;2011年09期
2 张祺丰;彭芸;李剑颖;刘s,
本文编号:1830383
本文链接:https://www.wllwen.com/yixuelunwen/yundongyixue/1830383.html