当前位置:主页 > 医学论文 > 特种医学论文 >

迭代重建技术(ASIR-V)在胸部CT检查中降低辐射剂量的初步研究

发布时间:2019-03-07 16:01
【摘要】:目的:探讨迭代重建技术(ASIR-V)在胸部Revolution CT检查中降低辐射剂量的应用价值。方法:随机收集120例拟行胸部RevolutionCT平扫的患者,将其随机分成4组,每组30例,分别在ASIR-V权重为0%(A组)、20%(B组)、40%(C组)、60%(D组)时扫描。扫描条件:噪声指数均设定为9,管电压为120KV,管电流为自动毫安(Smart mA 80-350),X线球管旋转速度0.5s,螺距0.992:1。患者检查前均测量身高及体重,并计算身高体重指数(body mass index,BMI)。记录每例患者扫描时的容积CT剂量指数(volume computed tomography dose index,CTDIvol)、剂量长度乘积(dose length product,DLP),并根据公式有效剂量(effective dose,ED)(mSv)=k(mSv·mGy-1·cm-1)×DLP(mGy·cm),计算ED值(k=0.019),对四组CT图像进行客观评价,测量其噪声(SDa)、肺动脉与竖脊肌的对比噪声比(CNRp)、肺动脉信噪比(SNRp)。并对四组图像肺窗及纵隔窗图像进行主观质量评分。然后采用单因素方差分析处理各组间的差异。结果:(1)随着ASIR-V权重增加,患者的有效剂量逐渐降低,两者呈负相关,四组数据间存在统计学差异(P0.05),A组有效剂量最大(5.13±0.48 mSv),D组有效剂量最小(1.97±0.50 mSv),剂量仅为A组的38.4%;(2)四组噪声之间存在统计学差异(P0.05),前三组噪声(SDa)随着ASIR-V权重的增加,数值逐渐减小,而D组噪声为四组中最大(8.81±1.91);(3)四组肺动脉与竖脊肌的对比噪声比(CNRp)存在统计学差异(P0.05),CNRp在C组(40%ASIR-V)最大(1.54±1.01),而在B组最小(0.998±1.07);(4)四组肺动脉信噪比(SNRp)在不同ASIR-V权重时也存在统计学差异(P0.05),SNRp在前三组呈递增趋势,C组最大(6.10±1.90),D组降为最低(4.84±1.60);(5)图像质量主观评价方面,四组主观评价分值之间存在统计学差异(P0.05),四组评分呈递减趋势,且前三组递减趋势比较微弱,D组评分明显降低,A组评分最高,值为14.2±1.03,D组评分最低,为8.7±1.12。结论:综合有效辐射剂量、图像质量的主客观指标,Revolution CT行胸部平扫时采用40%权重ASIR-V能有效降低辐射剂量,同时又能保证图像的质量。
[Abstract]:Objective: to evaluate the value of iterative reconstruction technique (ASIR-V) in reducing radiation dose in thoracic Revolution CT. Methods: one hundred and twenty patients were randomly divided into 4 groups (30 patients in each group) with ASIR-V of 0% (group A), 20% (group B), 40% (group C) and 60% (group D) when the ASIR-V weight was 0% (group A), 20% (group B), 40% (group C) and 60% (group D), respectively. Scanning conditions: noise figure is set to 9, tube voltage is 120kV, tube current is automatic milliampere (Smart mA 80kv 350), rotating speed of X-ray tube is 0.5s, pitch of screw is 0.992ml. Height and weight were measured before examination, and body height and body mass index (body mass index,BMI) were calculated. The volume CT dose index (volume computed tomography dose index,CTDIvol) and the dose length product (dose length product,DLP) of each patient were recorded at the time of scanning, and the effective dose of (effective dose, was calculated according to the formula. ED) (mSv) = k (mSv 路mGy-1 cm-1) 脳 DLP (mGy cm), calculated Ed value (k = 0.019). Four groups of CT images were evaluated objectively, and the noise ratio of (SDa), pulmonary artery to vertical spinal muscle was measured by comparing the noise to noise ratio of (SDa), pulmonary artery to vertical spinal muscle. Pulmonary artery signal-to-noise ratio (SNRp). The subjective quality of lung window and mediastinal window were evaluated. Then one-way ANOVA was used to deal with the differences among groups. Results: (1) with the increase of the weight of ASIR-V, the effective dose of the patients decreased gradually, and there was a negative correlation between the two groups. There was statistical difference among the four groups (P 0.05). The effective dose of), A group was the highest (5.13 卤0.48 mSv),). The effective dose in group D was only 38.4% of that in group A (1.97 卤0.50 mSv),). (2) there was statistical difference among the four groups of noise (P0.05). The noise (SDa) of the first three groups decreased gradually with the increase of ASIR-V weight, while the noise of group D was the highest among the four groups (8.81 卤1.91). (3) there was a significant difference in the noise ratio between the pulmonary artery and the vertical spine muscle in the four groups (P0.05), CNRp was the largest (1.54 卤1.01) in group C (40%ASIR-V), and the smallest (0.998 卤1.07) in group B; the noise ratio of (CNRp) was significantly higher in group C (40%ASIR-V) than in group B; (4) there was significant difference in the signal-to-noise ratio (SNR) (SNRp) of pulmonary artery between the four groups when the weight of ASIR-V was different (P0.05), SNRp increased in the first three groups, and the highest in group C (6.10 卤1.90), D) was the lowest (4.84 卤1.60); (5) in the subjective evaluation of image quality, there was a statistical difference among the four groups (P0.05), the scores of the four groups showed a decreasing trend, and the first three groups showed a weak decreasing trend, the score of Group D was significantly lower, and the score of Group A was the highest. The score of group D was the lowest (8.7 卤1.12). Conclusion: combining the effective radiation dose and the subjective and objective index of image quality, Revolution CT, 40% weighted ASIR-V can effectively reduce the radiation dose and guarantee the quality of the image during chest plain scan.
【学位授予单位】:苏州大学
【学位级别】:硕士
【学位授予年份】:2016
【分类号】:R816.4

【相似文献】

相关期刊论文 前10条

1 王海丹;;英国发布近5年的医疗辐射剂量报告[J];国外核新闻;2008年01期

2 黄桂玲;孙戈新;刘剑平;;心脏CT成像低辐射剂量技术的进展[J];中国老年学杂志;2010年14期

3 何雅静;郭丽敏;;降低放射给药辐射剂量[J];中国卫生质量管理;2011年03期

4 黄润玲;;广州市部分医院介入放射操作人员辐射剂量水平调查[J];华南预防医学;2011年04期

5 张红,张华东,黄文胜,桂正平,孙玲;重庆市辖区建筑材料成品表面辐射剂量调查[J];中华放射医学与防护杂志;2000年01期

6 张红,张华东,任在鸣,黄文胜,桂正平;辖区建筑材料成品表面辐射剂量调查[J];职业卫生与病伤;2000年01期

7 张翼,曲桂莲,张丹枫;照射野面积与透视脉冲频率对介入手术操作位辐射剂量的影响[J];中华放射医学与防护杂志;2001年06期

8 王宁;第三大医护中心病人辐射剂量校正[J];国外医学(放射医学核医学分册);2002年03期

9 尹艳,李红,文湘闽,李建成;燃煤电厂排放灰复田对公众所致的辐射剂量评价[J];职业卫生与病伤;2002年04期

10 张继勉;组合移动式集装箱检查系统辐射剂量的测量[J];中华放射医学与防护杂志;2003年06期

相关会议论文 前10条

1 来守永;;低辐射剂量摄影放在脑海中的第一位[A];2010中华医学会影像技术分会第十八次全国学术大会论文集[C];2010年

2 袁建华;;降低辐射剂量任重道不远[A];2013年浙江省放射学学术年会论文集[C];2013年

3 郑晖;;数字影像质量与如何控制受检者辐射剂量的思考[A];2009中华医学会影像技术分会第十七次全国学术大会论文集[C];2009年

4 郁鹏;付光新;王雷;燕树林;尉可道;刘澜涛;王建超;程玉玺;;介入诊疗中第一术者可控参数对其辐射剂量影响的测定与评价[A];2009中华医学会影像技术分会第十七次全国学术大会论文集[C];2009年

5 张福君;卓水清;陈林;张亮;赵明;刘健;李奎;郑庆生;王琰;;~(125)Ⅰ放射性粒子植入术后患者周围辐射剂量的监测与防护[A];第四届中国肿瘤学术大会暨第五届海峡两岸肿瘤学术会议论文集[C];2006年

6 张福君;卓水清;陈林;张亮;赵明;郑庆生;王琰;焦德超;;125I放射性粒子植入术后患者周围辐射剂量的监测与防护[A];中国(第七届)肿瘤微创治疗学术大会暨世界影像导引下肿瘤微创治疗学会成立筹备大会论文汇编[C];2011年

7 路春兰;;介入放射学在小儿先天性心脏病诊断治疗中的辐射剂量调查[A];2010中华医学会影像技术分会第十八次全国学术大会论文集[C];2010年

8 廖大富;;放射介入性检查工作人员之辐射剂量评估[A];2010中华医学会影像技术分会第十八次全国学术大会论文集[C];2010年

9 张翼;宋少娟;;基于数字化影像的病人辐射剂量监控方法[A];2009中华医学会影像技术分会第十七次全国学术大会论文集[C];2009年

10 赵永霞;黄桦;刘笑飞;;三种乳腺X线摄影系统影像质量和辐射剂量的比较[A];中华医学会第十八次全国放射学学术会议论文汇编[C];2011年

相关重要报纸文章 前10条

1 人民医院 綦维维 华东医院 王鸣鹏 江苏省人民医院 刘伟;优化扫描方案 合理控制辐射[N];中国医药报;2010年

2 武汉大学人民医院副研究员 罗照春;避免滥用CT 减少患癌几率[N];保健时报;2010年

3 记者 汤江峰;CT滥用:癌魔悄悄逼近[N];大众卫生报;2009年

4 本报记者 操秀英;核,其实就在身边[N];科技日报;2014年

5 记者 李学华;南非公布核电站辐射泄漏调查结果[N];科技日报;2010年

6 记者 卞晨光;儿童用CT扫描辐射剂量可能过高[N];科技日报;2010年

7 本报记者 赵家新 通讯员 彭健 沈益;监测辐射剂量,,把放射源管在“眼皮底下”[N];人民公安报;2014年

8 本报记者 衣晓峰 通讯员 冯宇曦;日常生活如何防辐射[N];中国中医药报;2011年

9 记者 张旭;GE推出世界首款宝石能谱CT[N];中国医药报;2009年

10 刘昆;莫让X线检查伤害儿童[N];大众卫生报;2009年

相关博士学位论文 前4条

1 刘洋;基于可视中国人体素数据的内辐射剂量研究[D];华中科技大学;2011年

2 孙文娟;基于数字人体的程式化模型和曲面模型构建及外辐射剂量研究[D];华中科技大学;2012年

3 张殿星;胸部数字化体层融合的辐射剂量研究与对肺内小结节探查的应用评价[D];山东大学;2012年

4 姚青山;非人类物种辐射剂量评估方法研究[D];中国原子能科学研究院;2006年

相关硕士学位论文 前10条

1 许静;CT、DSA设备辐射剂量的测量与研究[D];泰山医学院;2014年

2 秦好朴;婴幼儿胸部X线数字化高电压摄影技术的研究[D];河北医科大学;2015年

3 侯毅;宝石CT“双低”技术在冠脉成像中的临床应用[D];山西医科大学;2016年

4 蔡湘琴;卷荚相思种子~(60)Co-γ辐射诱变效应的初步研究[D];福建农林大学;2016年

5 吕志征;C-arm CT在神经介入手术中辐射剂量的研究[D];吉林大学;2016年

6 辛立旭;迭代重建技术(ASIR-V)在胸部CT检查中降低辐射剂量的初步研究[D];苏州大学;2016年

7 杨朝慧;低剂量64排螺旋CT冠状动脉成像扫描技术研究[D];山西医科大学;2014年

8 李勰;低辐射剂量CT肺动脉成像的临床应用研究[D];南方医科大学;2016年

9 汪海燕;~(125)I对兔喉组织放射性损伤及周围辐射剂量的实验研究[D];暨南大学;2012年

10 师勤莹;心血管介入诊疗术者站立区域辐射剂量的测量与分析[D];山西医科大学;2014年



本文编号:2436250

资料下载
论文发表

本文链接:https://www.wllwen.com/yixuelunwen/yundongyixue/2436250.html


Copyright(c)文论论文网All Rights Reserved | 网站地图 |

版权申明:资料由用户b075c***提供,本站仅收录摘要或目录,作者需要删除请E-mail邮箱bigeng88@qq.com