显微拉曼光谱法定量检测碳氧血红蛋白饱和度
发布时间:2018-04-03 16:41
本文选题:拉曼光谱 切入点:碳氧血红蛋白 出处:《科学通报》2012年14期
【摘要】:采用显微拉曼光谱技术对不同饱和度碳氧血红蛋白(HbCO)进行测定,通过偏最小二乘法(PLS)和间隔偏最小二乘法(IPLS)建立模型,得出线性回归方程.结果表明,经过筛选变量信息所建立的间隔偏最小二乘模型(IPLS),当血红蛋白系列样品中碳氧血红蛋白含量为0~100%时,线性相关系数的平方为0.99,检测限为5.37%,完全能满足临床和法医学上CO中毒程度的检测.此方法具有快速、简便、直接、对样品无损等优点,有望成为检测碳氧血红蛋白饱和度的新方法.
[Abstract]:The different saturation carbon oxygen hemoglobin (HbCO-HbCO) was measured by micro-Raman spectroscopy. The linear regression equation was obtained by using partial least squares (PLS) and interval partial least squares (IPLSs).The results showed that the interval partial least squares model was established by screening the information of variables, when the content of carboxyhemoglobin in hemoglobin series samples ranged from 0 to 100.The square of linear correlation coefficient is 0.99 and the detection limit is 5.37, which can fully meet the clinical and forensic detection of CO poisoning.This method has the advantages of rapidity, simplicity, directness and nondestructive to the sample, so it is expected to be a new method for the determination of carboxyhemoglobin saturation.
【作者单位】: 华南师范大学生物光子学研究院光子中医学实验室 激光生命科学教育部重点实验室;赣南师范学院物理与电子信息学院;
【基金】:国家自然科学基金(60878063) 广东省中医药局项目(2008233)资助
【分类号】:D919.4
【参考文献】
相关期刊论文 前1条
1 姜宴,叶永红,张宜凡,吴文庆;血中碳氧血红蛋白饱和度测定影响因素的研究[J];法医学杂志;2003年02期
【共引文献】
相关期刊论文 前7条
1 姜宴,叶永红,张宜凡,吴文庆;血中碳氧血红蛋白饱和度测定影响因素的研究[J];法医学杂志;2003年02期
2 刘冬娴,胡启芬,王彦华,谭家镒;血液中碳氧血红蛋白饱和度的双波长分光光度简便测定法[J];光谱实验室;2001年01期
3 孟耀勇,,刘颂豪;三链DNA分子poly(dT)·poly(dA)·poly(dT)的低频振动模式[J];光学学报;1994年07期
4 孟耀勇,刘颂豪;三螺旋DNA分子久期方程的约化[J];华南师范大学学报(自然科学版);1993年02期
5 白璐,廖林川,颜有仪;高效液相色谱法检测肝组织中马钱子碱和士的宁[J];刑事技术;2005年04期
6 尉志文;
本文编号:1706033