福州市医用诊断X射线机摄影所致受检者照射剂量研究
[Abstract]:Objective to investigate the (ESD) level of incident surface dose in patients with medical diagnostic X-ray camera in Fuzhou. Methods by cluster sampling, 52 medical diagnostic X-ray cameras were selected from 38 hospitals in Fuzhou in 2014, which were qualified for image quality control. The position of the radiographs included head, chest, lumbar vertebrae and lateral position of thoracic vertebrae. The standard water motif was used to simulate the adult subjects, and the thermoluminescence meter was placed in the central position of the irradiation field of the standard water mode. The ESD was detected by exposure under normal clinical irradiation conditions. Results the ESD of the subjects examined by X-ray radiography was 0.35 ~ 17.90 m Gy, and the position of the chest was the lowest (0.35 卤0.03) m Gy,). The lateral position of the lumbar vertebrae was the highest. For (17.90 卤1.26) m Gy.8 position X-ray camera, the ESD of the subjects were lower than that of GB 16348-2010. The standard for radiological health protection of medical X-ray diagnostic examinees was lower than that recommended by GB 16348-2010. Dose guidance level of X-ray radiography in subjects. The direct digital X-ray imaging system (DR) and the computer X-ray imaging system (CR) caused ESD in the head, lumbar vertebrae and thoracic vertebrae were lower than those in the normal X-ray machine (P0.05). There was no significant difference in the positive position and lateral position of the head, lumbar vertebrae and thoracic vertebrae caused by DR and CR (P0.05). There was no significant difference in ESD between the above three X-ray cameras in the positive and lateral chest position (P0.05). Conclusion the ESD of the patients with DR and CR radiography is lower than that of the ordinary X-ray machines. The popularization of DR and CR is helpful to reduce the dose of radiation.
【作者单位】: 福建医科大学公共卫生学院;福州市疾病预防控制中心;
【分类号】:R144
【相似文献】
相关期刊论文 前10条
1 米砚祥;;波拉胶片在工业射线摄影上的应用[J];感光材料;1979年05期
2 蔡景猷;;射线摄影影象增强法[J];感光材料;1983年01期
3 肖石英;陈顺妹;;一种新的弯曲射线摄影方法[J];世界采矿快报;1993年08期
4 李连波,杨珂,何顺升,于夕荣;X射线摄影中防护剂量检测方法的探讨[J];中华放射医学与防护杂志;2000年03期
5 闫革彬;孙立伟;;床旁X射线摄影防护现状与对策[J];中国辐射卫生;2009年04期
6 有田正史;卢林松;;沉积物软X射线摄影的样品制备[J];海洋地质译丛;1983年06期
7 韩相明;;采用“T颗粒”技术的新型射线摄影胶片[J];感光材料;1985年01期
8 王鹏程,张富利,李士俊;骨盆数字化X射线摄影曝光剂量的优化研究[J];中华放射医学与防护杂志;2005年05期
9 裘朱熙;;柯达X-OMAT AR胶片[J];化工新型材料;1981年11期
10 于柏松;张洪欣;张松;;鞍山地区医用诊断X射线摄影设备质量控制检测结果分析[J];工业卫生与职业病;2014年03期
相关会议论文 前3条
1 舒震宇;;数字化X线摄影的应用及发展[A];2005年浙江省放射学学术年会论文汇编[C];2005年
2 唐馗阳;杨方;乔国庆;欧陕兴;;CR系统的临床应用研究与对策[A];中国人民解放军医学会第九届放射诊疗专业学会论文集[C];2005年
3 汪新柱;陈晓飞;贾有福;;儿童肘关节DR卧位摄影法[A];2010中华医学会影像技术分会第十八次全国学术大会论文集[C];2010年
相关重要报纸文章 前4条
1 本报记者 诸玲珍;X射线摄影系统掀起数字潮[N];中国电子报;2000年
2 ;直接数字化X射线摄影技术[N];中国电子报;2000年
3 记者 何永晋;乌克兰学者提出管道石油计算新方法[N];科技日报;2002年
4 健康时报记者 赵安平;少儿体检为何禁止做胸透?[N];健康时报;2006年
相关博士学位论文 前2条
1 刘建军;胸部数字X射线摄影所致中国成人受检者剂量研究[D];中国疾病预防控制中心;2008年
2 刘海宽;X射线诊断所致受检者剂量的估算模式研究[D];复旦大学;2009年
相关硕士学位论文 前3条
1 马天俊;数字化X射线摄影所致受检者辐射剂量估算研究[D];复旦大学;2013年
2 李鹏;常见类型医用X射线摄影的剂量指导水平研究[D];复旦大学;2013年
3 董泽成;PE公司数字X射线摄影设备(DR)营销策略研究[D];苏州大学;2014年
,本文编号:2345982
本文链接:https://www.wllwen.com/yixuelunwen/yufangyixuelunwen/2345982.html