脊柱动态固定的剪切特性与载荷分享的生物力学研究

发布时间:2018-03-18 00:38

  本文选题:脊柱动态固定 切入点:剪切 出处:《南方医科大学》2017年博士论文 论文类型:学位论文


【摘要】:目的:1.探讨多节段退行性滑脱生物力学模型的建立,比较L4-L5及L5-S1节段的剪切特性。2.评价腰骶椎后路PEEK棒、滑动棒、转动钉三种动态固定的剪切稳定性。3.评价胸腰椎前路PEEK棒、滑动棒、转动钉三种动态固定的稳定性与载荷分享。方法:1.对8例人体标本的L4-L5及L5-S1节段在300N轴向压缩力下分别于两间隙施加-50N~250N前后方向剪切力。测试完整、小关节损伤及椎间盘损伤三种状态下节段向前的剪切位移,计算剪切刚度。2.采用8例L4-L5节段退行性滑脱模型,在有或无椎间融合器(Cage)植入两种情形下分别测量后路刚性钉棒及PEEK棒、滑动棒、转动钉固定下向前的剪切位移。剪切加载方式同上。3.对8例人体T11-L3节段标本分别在屈伸、侧弯及旋转方向施加5Nm的纯力偶矩,进一步在300N跟随力加载下在屈伸方向施加5Nm纯力偶矩。完整状态测试后行前路L1椎体次全切除,采用椎间模拟钛网(含微力传感器)及T12-L2节段前路钉棒进行内固定重建。分别在椎间钛网全长及短缩2mm(模拟植骨沉降)两种情形下对前路刚性及PEEK棒、滑动棒、转动钉固定的节段运动范围(range of motion,ROM)及椎间载荷等进行测量。结果:1.L4-L5及L5-S1节段椎间盘损伤状态的前剪切位移分别为1.4mm及1.2mm,均较完整状态位移显著增加(P0.05)。2.完整L5-S1节段的前剪切刚度较L4-L5节段显著增加(P0.05),在小关节及椎间盘损伤状态下两节段的前剪切刚度相当。3.后路PEEK棒及转动钉固定的前剪切位移较刚性固定显著增加(P0.05),后路滑动棒固定的位移与之相当。4.椎间Cage植入后所有内固定下的前剪切位移均较Cage植入前减少,动态固定位移减少更明显。5.无论椎间钛网全长或短缩,与前路刚性固定比较,PEEK棒及转动钉固定在各个方向上的ROM均不同程度增加,滑动棒固定的ROM在钛网短缩的前屈方向(有跟随力)减少,在其余方向上与之相当或略增加。6.无论椎间钛网全长或短缩,300N跟随力下三种前路动态固定的椎间载荷值在中立位及屈伸位均大于刚性固定。结论:1.L4-L5及L5-S1节段可同时建立I°退行性滑脱模型。2.完整L5-S1节段比L4-L5节段能更好的抵抗剪切滑脱,但两节段在损伤状态抵抗剪切滑脱的能力相当。3.腰骶椎后路PEEK棒、转动钉固定的剪切稳定性弱于刚性固定,滑动棒固定的剪切稳定性与之相当。4.椎间Cage植入能增强所有后路固定的剪切稳定性,但Cage植入对动态固定的影响要大于刚性固定。5.胸腰椎前路PEEK棒及转动钉固定的稳定性整体上弱于刚性固定,滑动棒固定稳定性与刚性固定相当。6.胸腰椎前路PEEK棒、滑动棒及转动钉固定较刚性固定有更好的椎间植骨载荷分享。
[Abstract]:Objective 1. To establish a biomechanical model of multilevel degenerative spondylolisthesis and compare the shear characteristics of L4-L5 and L5-S1 segments. Stability and load sharing of three kinds of dynamic fixation of sliding rod and rotating nail. Methods: 1. The shear forces of L4-L5 and L5-S1 segments in 8 human specimens were applied to two clearances before and after -50N 250N respectively under 300N axial compression force. The forward shear displacement of small joint injury and intervertebral disc injury was calculated. 2. 8 cases of L4-L5 degenerative slippage model were used. The shear displacement of posterior rigid screw rod and PEEK rod, sliding rod and rotating screw fixation were measured in two cases with or without CageCage. The shear loading mode was as follows: 3. For 8 cases of human T11-L3 segment specimens, they were flexion and extension, respectively. The pure force couple moment of 5Nm was applied in the lateral bend and rotation direction, and the 5Nm pure force moment was applied in the flexion and extension direction under 300N following force. After the complete state test, the anterior L1 vertebra was subtotal excised. The intervertebral titanium mesh (including micro-force sensor) and T12-L2 segment anterior screw rod were used for internal fixation reconstruction. The anterior rigid and PEEK rod, sliding rod, were obtained under the condition of full length of intervertebral titanium mesh and short reduction of 2mm (simulated bone graft subsidence), respectively. The range of motion and the intervertebral load were measured. Results the anterior shear displacement of 1. 1. L4-L5 and L5-S1 segments were 1.4 mm and 1. 2 mm, respectively, which increased P0. 05%. 2. The intact L5-S1 segment. The anterior shear stiffness of L4-L5 segment was significantly higher than that of L4-L5 segment, and the anterior shear stiffness of the two segments was about the same as that of L4-L5 segment under the condition of facet joint and disc injury. The anterior shear displacement of posterior PEEK rod and rotation screw fixation was significantly higher than that of rigid fixation, and the posterior sliding was significantly higher than that of L4-L5 segment. The displacement of rod fixation was the same. 4. After intervertebral Cage implantation, the anterior shear displacement of all internal fixation was less than that of Cage. The reduction of dynamic fixation displacement was more obvious. Regardless of the total length or short contraction of the intervertebral titanium mesh, the ROM of peek rod and rotating nail were increased in different degrees compared with the anterior rigid fixation. The ROM fixed by the sliding rod is reduced in the short flexion direction (with following force) of the titanium mesh. In other directions, the intervertebral load of the three kinds of dynamic anterior fixation is larger than that of rigid fixation in neutral position and flexion and extension position, regardless of the total length of the intervertebral titanium mesh or the short contraction 300N following force. Conclusion: 1. The segments of L4-L5 and L5-S1 can be the same as those of rigid fixation. The complete L5-S1 segment was more resistant to shearing slippage than L4-L5 segment. However, the ability of two segments to resist shear slippage in the injured state is equal. 3. The shear stability of rotation nail fixation is weaker than that of rigid fixation in lumbosacral posterior approach PEEK rod. The shear stability of sliding rod fixation is equivalent to that of slide-rod fixation. 4. Intervertebral Cage implantation enhances the shear stability of all posterior fixations. But the effect of Cage implantation on dynamic fixation was greater than that on rigid fixation. The stability of thoracolumbar anterior PEEK rod and rotation screw fixation was weaker than that of rigid fixation, and the stability of sliding rod fixation was equal to that of rigid fixation. Sliding rod and screw fixation are better than rigid fixation in load sharing of intervertebral bone graft.
【学位授予单位】:南方医科大学
【学位级别】:博士
【学位授予年份】:2017
【分类号】:R681.5

【相似文献】

相关期刊论文 前10条

1 本刊编辑部;生物力学研究文稿的写作须知[J];中国骨伤;2002年10期

2 张松林;;第4届世界生物力学大会[J];体育科学;2003年01期

3 ;第三届《医用生物力学》编委会[J];医用生物力学;2003年03期

4 许燕,樊瑜波;法医生物力学研究进展[J];生物医学工程学杂志;2004年01期

5 本刊编辑部;生物力学研究文稿的写作须知[J];中国骨伤;2004年11期

6 ;《医用生物力学》更改为双月刊启事[J];国际心血管病杂志;2008年01期

7 秦岭;;生物力学——一门跨学科的学科[J];医用生物力学;2008年02期

8 龙勉;;关于国家自然科学基金生物力学学科优先资助领域的思考[J];医用生物力学;2009年S1期

9 姜宗来;;我国生物力学研究发展三十年[J];医用生物力学;2010年S1期

10 樊瑜波;;生物力学在医疗器械设计和跳伞着陆防护中的应用[J];医用生物力学;2010年S1期

相关会议论文 前10条

1 姜宗来;;我国生物力学学科发展的思考[A];第九届全国生物力学学术会议论文汇编[C];2009年

2 杨辰;曲峰;;裸足运动的生物力学研究综述[A];第十五届全国运动生物力学学术交流大会(CABS2012)论文摘要汇编[C];2012年

3 樊瑜波;;医疗器械及人工器官的生物力学设计[A];第九届全国生物力学学术会议论文汇编[C];2009年

4 成自龙;;人体冲击生物力学研究进展[A];第四届全国人—机—环境系统工程学术会议论文集[C];1999年

5 迪丽娜·马合木提;穆合塔尔·沙德尔;哈木拉提·吾甫尔;;维医沙疗中的生物力学问题之探讨[A];“力学2000”学术大会论文集[C];2000年

6 孟和;钟红刚;;生物力学与中西医结合骨科的发展[A];第九次全国中西医结合创伤骨科学术大会论文汇编[C];2001年

7 杨桂通;;探索人体奥秘的古老故事——生物力学孕育期一瞥[A];力学史与方法论论文集[C];2003年

8 杨桂通;;探索人体奥秘的古老故事——生物力学孕育期一瞥[A];力学史与方法论论文集[C];2003年

9 龚璐璐;丁祖泉;高建新;冷晔;;数字散斑相关法及其在医用生物力学领域的应用[A];第八届全国生物力学学术会议论文集[C];2006年

10 刘振田;吴小涛;;生物力学中多频振动技术治疗的现状和展望[A];中国生物医学工程学会第六次会员代表大会暨学术会议论文摘要汇编[C];2004年

相关重要报纸文章 前4条

1 汪雪姣 林岚;辛勤耕耘在生物力学园圃中[N];科技日报;2004年

2 黄盈盈;走出困境 向高尖端领域发展[N];福建科技报;2008年

3 王美宽;运动品牌重视核心竞争力塑造[N];中国服饰报;2008年

4 陈云飞;上海岳阳医院 推拿项目获地方科技一等奖[N];中国中医药报;2010年

相关博士学位论文 前10条

1 孟祥杰;基于肌肉骨骼生物力学载荷的驾驶人腰部舒适性研究[D];清华大学;2015年

2 麻恒源;唇裂唇鼻肌肉系统三维解剖学研究与唇裂功能性修复的有限元生物力学模似[D];北京协和医学院;2016年

3 赵云飞;力学性能优化的青少年特发性脊柱侧凸三维有限元模型建立及置钉策略的生物力学和临床回顾研究[D];第二军医大学;2017年

4 周若舟;脊柱动态固定的剪切特性与载荷分享的生物力学研究[D];南方医科大学;2017年

5 许硕贵;动态记忆应力的生物力学研究及临床意义探讨[D];第二军医大学;2001年

6 苏佳灿;髋臼三维记忆内固定系统治疗髋臼骨折记忆生物力学研究——骨盆、髋臼三维模型仿真、力学模拟与有限元分析[D];第二军医大学;2004年

7 张美珍;非接触性前交叉韧带损伤危险因素的生物力学研究[D];北京体育大学;2012年

8 赵文志;活骨组织应力与重建适应实验及其生物模型研究[D];大连理工大学;2004年

9 连志强;骨骼材料与结构的性能仿真及重建模拟研究[D];大连理工大学;2010年

10 陈强;挥鞭样损伤的生物力学和临床研究[D];第二军医大学;2005年

相关硕士学位论文 前10条

1 王晓猛;三种sMCL和POL重建方法的生物力学对比性研究[D];河北医科大学;2015年

2 郭伟明;腺病毒介导TGFβ1基因体外转染对自体Hamstring腱重建兔ACL的生物力学及影像学影响[D];南华大学;2015年

3 冯国坤;油茶果与花苞生物力学测量系统设计与研究[D];江西农业大学;2015年

4 赵鹏飞;正常男性腰椎三维有限元模型的建立并证明其在所给力矩载荷下的有效性[D];大连医科大学;2015年

5 林田;基于生物力学的鞋舒适性评价和设计[D];福州大学;2014年

6 吴凯;基于损伤生物力学的颈部精细化建模及在低速后碰撞中的应用[D];华南理工大学;2016年

7 韩硕;有限元分析在腰椎间盘生物力学中的应用研究[D];河北医科大学;2016年

8 聂兵;正常人耳廓生物力学基本特征的研究[D];第二军医大学;2016年

9 田东牧;骨干假体与辅助钛板联合重建肱骨干缺损的生物力学研究[D];天津理工大学;2017年

10 纪祥;拇指掌指关节侧副韧带损伤的生物力学研究及临床应用[D];吉林大学;2007年



本文编号:1627247

资料下载
论文发表

本文链接:https://www.wllwen.com/shoufeilunwen/yxlbs/1627247.html


Copyright(c)文论论文网All Rights Reserved | 网站地图 |

版权申明:资料由用户81a99***提供,本站仅收录摘要或目录,作者需要删除请E-mail邮箱bigeng88@qq.com